MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE¿ IMPLANT 3.75 X 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE¿ IMPLANT 3.75 X 11.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number OSS311
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A1: patient identifier: unknown / not provided.
 
Event or Problem Description
The doctor reports that the dental implant located in position number #15 failed because it fractured at the body of the implant and was removed.The doctor reports that the procedure was not concluded by placing another implant.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D4: additional device information.D9: device availability.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) oss311, (osseotite® implant 3.75 x 11.5mm) for evaluation.Visual evaluation of the as returned device identified the implant with signs of use, and was attached to a crown.The implant was fractured at the body region.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number 2017040406.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2017040406 for similar events and no other complaint was identified.Review completed utilizing keywords: fracture implant.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are parafunctional habits/patient factors, or customer error.Therefore, based on the available information, a device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE¿ IMPLANT 3.75 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key20902684
Report Number0001038806-2024-02890
Device Sequence Number17394732
Product Code DZE
UDI-Device Identifier00844868013112
UDI-Public(01)00844868013112(17)220421(10)2017040406
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K063286
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/21/2022
Device Catalogue NumberOSS311
Device Lot Number2017040406
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/11/2024
Supplement Report FDA Received Date04/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/21/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
Patient Weight78 KG
Patient EthnicityNon Hispanic
-
-