Catalog Number SPD2-US-060-320
Medical Device Problem Codes Entrapment of Device (1212); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a procedure to treat a calcified lesion in the right proximal mid distal superficial femoral artery with 80% stenosis and an 80mm lesion length, there were device-related issues involving the turbohawk plus ls and the spider fx devices.Minimal vessel tortuosity and severe vessel calcification were reported.Artery diameter reported as 7mm.The procedure involved the use of a non-medtronic 7f sheath, and pre- and post-dilation with 3mm and 6mm nanocross devices, respectively.The physician made several passes with turbohawk plus.Upon cleaning and reinserting the device into the body, the turbohawk plus nose became bound to the spider fx wire, making it extremely difficult to advance.Severe resistance was encountered for each device, and excessive force was used during the procedure.Removal difficulties were reported.The physician used clamps and a technician's aid to forcefully remove the hawk from the sheath and pull it off the wire.The devices were removed successfully in tandem without injury, and the procedure was completed using new medtronic devices, specifically a turbohawk plus ls and a 6mm spider fx, with no further issues.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIDER FX
Common Device Name
TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key20903202
Report Number2183870-2024-00463
Device Sequence Number5006066
Product Code NTE
UDI-Device Identifier00821684068366
UDI-Public00821684068366
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111010
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/09/2026
Device Catalogue NumberSPD2-US-060-320
Device Lot NumberB757032
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2024
Supplement Date Received by Manufacturer02/13/2025
Initial Report FDA Received Date12/11/2024
Supplement Report FDA Received Date02/18/2025
Date Device Manufactured07/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PRODUCT ID THP-LS (0012096423)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
Patient Weight75 KG
-
-