|
D10 concomitant product: sigphandle, sig power sigphandle handle, (serial # (b)(6) sigpshell, sig power sigpshell control shell, (lot #unknown) sigadaptstnd, sig power sigadaptstnd linear adapter, (serial # (b)(6)).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
According to the reporter, during a pulmonary resection, an abnormal sound had sounded from the device, and it fired partially.It was noted that the power handle was restarted to confirm the issue.The product was replaced to resolve the issue.There was no patient injury.
|