BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-35 4/150/130; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
|
Back to Search Results |
|
| Model Number |
383281 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The passeo-35 balloon system was selected for pre-dilatation of a moderately calcified lesion in a moderately tortuous proximal sfa, but the balloon was leaking during inflation.Another balloon was used to finish the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed that the balloon has been inflated and was deflated in the as-returned state.Upon inflation a fine jet of water was seen to emerge from the distal part of the balloon.Microscopic inspection revealed a small pinhole in the balloon surface about 23 mm proximal to the distal x-ray marker.Scratches were observed in close vicinity of the pinhole.It seems likely that the damage of the balloon surface was caused by a hard, sharp-edged object such as e.G., anatomical structure.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each device is tested for air tightness by means of a pressure test and a helium leak test.We can therefore confirm that the device was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy (i.E., calcified target lesion).
|
| |
|
Search Alerts/Recalls
|
|
|