• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-35 4/150/130; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-35 4/150/130; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 383281
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2024
Type of Reportable Event Malfunction
Event or Problem Description
The passeo-35 balloon system was selected for pre-dilatation of a moderately calcified lesion in a moderately tortuous proximal sfa, but the balloon was leaking during inflation.Another balloon was used to finish the procedure.
 
Additional Manufacturer Narrative
The returned device was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed that the balloon has been inflated and was deflated in the as-returned state.Upon inflation a fine jet of water was seen to emerge from the distal part of the balloon.Microscopic inspection revealed a small pinhole in the balloon surface about 23 mm proximal to the distal x-ray marker.Scratches were observed in close vicinity of the pinhole.It seems likely that the damage of the balloon surface was caused by a hard, sharp-edged object such as e.G., anatomical structure.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each device is tested for air tightness by means of a pressure test and a helium leak test.We can therefore confirm that the device was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy (i.E., calcified target lesion).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-35 4/150/130
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key20906050
Report Number1028232-2024-06349
Device Sequence Number9133022
Product Code LIT
UDI-Device Identifier07640130434639
UDI-Public(01)07640130434639(17)2612
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K142379
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number383281
Device Catalogue Number383281
Device Lot Number11230237
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2024
Supplement Date Received by Manufacturer01/06/2025
Initial Report FDA Received Date12/11/2024
Supplement Report FDA Received Date01/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-