Model Number WA2T43WA
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the subject device had a lens cracked.There were no reports of patient harm.
 
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: d2a - common device name: changed "wa2t43wa" with "rigid endoscope telescope".The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction due to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE "OES ELITE", 4 MM, 30 °, WIDE-ANGLE, HD, AUTOCLAVABLE
Common Device Name
RIGID ENDOSCOPE TELESCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrase 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrase 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20906787
Report Number9610773-2024-33842
Device Sequence Number4120618
Product Code FGC
UDI-Device Identifier04042761077569
UDI-Public04042761077569
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K151191
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 05/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA2T43WA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/12/2024
Supplement Date Received by Manufacturer05/02/2025
Initial Report FDA Received Date12/11/2024
Supplement Report FDA Received Date05/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/11/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-