• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI ATTAIN ABILITY PLUS MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ATTAIN ABILITY PLUS MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 429688
Medical Device Problem Code High Capture Threshold (3266)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that erroneously high left ventricular capture management thresholds were noted.The left ventricular (lv) lead and cardiac resynchronization therapy pacemaker (crt-p) remain in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: w1tr01 crt-p implanted: on (b)(6) 2023.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that erroneously high left ventricular capture management thresholds were noted.The left ventricular (lv) lead and cardiac resynchronization therapy pacemaker (crt-p) remain in use.No patient complications have been reported as a result of this event.It was further reported that the lv lead and crt-p were reprogrammed to turn of the device feature that automatically monitors pacing thresholds at periodic intervals.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN ABILITY PLUS MRI SURESCAN
Common Device Name
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key20907124
Report Number2649622-2024-33535
Device Sequence Number9181073
Product Code OJX
UDI-Device Identifier00643169601796
UDI-Public00643169601796
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P080006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/19/2018
Device Model Number429688
Device Catalogue Number429688
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2024
Supplement Date Received by Manufacturer03/28/2025
Initial Report FDA Received Date12/11/2024
Supplement Report FDA Received Date04/10/2025
Date Device Manufactured10/10/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-52 LEAD, 5076-58 LEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexMale
-
-