COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 18865
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/03/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a tracheal tube experienced a connector break at the tracheal tube interface, indicating physical damage.The issue occurred within a hospital setting.No patient was associated with this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3 and h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the 15mm connector was broken.A dimensional test was performed on the connector: the inner diameter of the distal end was within specification, the outer diameter of the distal end was within specification, and no bubbles were observed in the breakage damage.It was reported that the a tracheal tube experienced a connector break at the tracheal tube interface, indicating physical damage.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: if shortening of the endotracheal tube by cutting is considered, the tube should be cut and the connector reinserted prior to intubation.Tubes with 15mm connectors that cannot be removed with reasonable manipulation are not suitable for cutting.Always assure the connector is firmly seated in both the tracheal tube and the breathing circuit to prevent disconnection during use.The 15mm connector is seated so that it can be removed with effort if pre-cutting of the tube is desired.Follow the suggested directions for use to evaluate the tube and connector for suitability if pre-cutting is considered.Always assure the connector is firmly seated in both the tracheal tube and the breathing circuit to prevent disconnection during use.Non-standard dimensioning of some connectors on ventilatory or anesthesia equipment may make secure mating with the tracheal tube 15mm connector difficult.Use only with equipment having standard 15mm connectors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20909932
Report Number9681384-2024-00232
Device Sequence Number8599592
Product Code BTR
UDI-Device Identifier40884521032515
UDI-Public40884521032515
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K090352
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number18865
Device Catalogue Number18865
Device Lot Number22K1306JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2024
Supplement Date Received by Manufacturer02/20/2025
Initial Report FDA Received Date12/11/2024
Supplement Report FDA Received Date02/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-