• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG IPERIA 7 DR-T DF-1 PROMRI; ICD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG IPERIA 7 DR-T DF-1 PROMRI; ICD Back to Search Results
Model Number 392409
Medical Device Problem Code Premature End-of-Life Indicator (1480)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/21/2024
Type of Reportable Event Malfunction
Event or Problem Description
Device is eos with unexpected battery behavior.No adverse patient events were reported.The device was explanted.
 
Additional Manufacturer Narrative
We received your event description for the device and would like to thank you for supporting our post-market surveillance.As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.
 
Additional Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were reviewed.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.Upon receipt, the icd was subjected to an electrical analysis, revealing that the device could not be interrogated.The visual inspection of the device showed no anomalies.An unexpected battery voltage trend and a premature battery depletion could be confirmed during analysis.Please note, this icd is affected by the field safety corrective action, bio-lqc, initiated in march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPERIA 7 DR-T DF-1 PROMRI
Common Device Name
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key20910015
Report Number1028232-2024-06355
Device Sequence Number7824211
Product Code MRM
UDI-Device Identifier04035479128777
UDI-Public(01)04035479128777(17)200331
Combination Product (Y/N)N
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2020
Device Model Number392409
Device Catalogue Number392409
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2024
Supplement Date Received by Manufacturer12/16/2024
03/03/2025
Initial Report FDA Received Date12/11/2024
Supplement Report FDA Received Date12/16/2024
03/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/17/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age81 YR
Patient SexMale
-
-