| Catalog Number |
8065752181 |
| Medical Device Problem Code |
Decrease in Suction (1146)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A physician reported that the cumulative dissipated energy (cde) value did not increase and remained at zero during cataract surgery (with intraocular lenses implant).The surgery was completed after replacing the product with another one.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in d.9., h.3., h.6.And h.11.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned sample was visually inspected and functionally tested.Both irrigation and aspiration luers were intact.Flare shape was observed in the blue socket fitting at the aspiration tubing insertion end.The aspiration and irrigation free flow met specification.The root cause of the customer's complaint could not be established since the returned sample met specifications.No contributing factors could be identified that could cause the reported complaint.After the investigation of this complaint, it has been determined that this sample met specifications.Therefore, no action will be taken at this time.Based on our current tracking, there are no adverse trends for this reported complaint.Quality assurance will continue to monitor customer complaints via the complaint review meetings and will take action for any future occurrences as is deemed necessary.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|