• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DITRON DENTAL LTD ULTIMATE PRECISION IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DITRON DENTAL LTD ULTIMATE PRECISION IMPLANT; DENTAL IMPLANT Back to Search Results
Model Number ULT
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Circumstances are not clear due to the lack of provided information.To date, no information has indicated that the device or the way it was used has contributed to the described event.In addition, no signals were found indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to osseointegration problem, and escaping the available detections has been assessed and concluded to be unlikely in high level of confidence.If additional information is received indicating otherwise, a follow-up report will be submitted.
 
Event or Problem Description
Failure to osseointegrate, the implant fell down.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTIMATE PRECISION IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
DITRON DENTAL LTD
2 haofe st.
southern industrial zone
ashkelon, israel 78780 02
IS  7878002
Manufacturer (Section G)
DITRON DENTAL LTD
2 haofe st.
southern industrial zone
ashkelon, israel 78780 02
IS   7878002
Manufacturer Contact
2 haofe st.
southern industrial zone
ashkelon, israel 78780-02
MDR Report Key20910797
Report Number3010948854-2022-00363
Device Sequence Number12709851
Product Code DZE
UDI-Device Identifier07290109280646
UDI-Public07290109280646
Combination Product (Y/N)N
PMA/510(K) Number
K161497
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Remedial Action Other
Type of Report Initial
Report Date (Section B) 12/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberULT
Device Catalogue Number11420
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2022
Initial Report FDA Received Date12/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age89 YR
Patient SexFemale
-
-