| Catalog Number |
INT411 |
| Medical Device Problem Codes |
Failure to Osseointegrate (1863); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Sinus Perforation (2277)
|
| Date of Event |
06/19/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem g4: premarket identification: corrected g6: type of report h2: follow up type h10: additional narrative.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b2: outcomes attributed to adverse event b4: date of this report b5: describe event or problem g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h10: additional narrative zimvie received one (1) int411, (osseotite tapered certain implant 4 x 11.5mm) for evaluation.Visual evaluation was performed, the implant had signs of use with markings from removal.The implant had bone debris on its internal and external threads.No damage or signs of malfunction that would have contributed to the event.Measurements match drawing.Device history record (dhr) review was completed for the subject lot number 2021090938.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2021090938 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : non integration : perforated sinus.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was inadequate treatment planning and patient factors.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable with all the available information provided.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Event or Problem Description
|
|
The doctor reports that the dental implant located in dental position number #36 failed to integrate due to sinus perforation and was removed.The doctor reports in the per that the procedure was not concluded by placing another implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A1: patient identifier: unknown / not provided.A2: age at time of the event: unknown / not provided.A4: patient weight: unknown / not provided.
|
| |
|