• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE S; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE S; IMPLANTABLE LEAD Back to Search Results
Model Number 0262
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) triggered an alert due to an out of range right ventricular (rv) lead pacing impedance measurement however, no value was displayed.Technical services (ts) discussed the 24-hour clock and informed pacing lead impedances have measured an average of 1,900 ohms.Normal range for this lead is 1,800 ohms.Ts reviewed the logbook and there is no evidence of noise.Ts recommended an in-clinic interrogation to reset the alert and to consider raising the upper limit to 2,250 ohms if clinically appropriate.Additionally, isometrics was recommended.At this time, the system remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE S
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key20914268
Report Number2124215-2024-78638
Device Sequence Number12707361
Product Code NVY
UDI-Device Identifier00802526601828
UDI-Public00802526601828
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/25/2024
Device Model Number0262
Device Catalogue Number0262
Device Lot Number100470
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2024
Initial Report FDA Received Date12/12/2024
Date Device Manufactured01/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexFemale
-
-