• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRHYTHM TECHNOLOGIES, INC ZIO XT; RECORDER, MAGNETIC TAPE, MEDICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRHYTHM TECHNOLOGIES, INC ZIO XT; RECORDER, MAGNETIC TAPE, MEDICAL Back to Search Results
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Anaphylactoid / Anaphylaxis (2218)
Date of Event 11/13/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The patient presented with signs of an allergic reaction allegedly caused by the zio xt to their healthcare provider.The patient experienced itchy red marks on their chest.As a result, the device was removed, and the patient was prescribed a steroid cream-triamcinolone to apply to the affected area.
 
Additional Manufacturer Narrative
A review of the complaint revealed that the patient wore the xt for 4 of the 14-day prescribed wear period.The patient has a history of reactions to medical adhesives and sensitive skin.The exact cause of the reported event cannot be determined.However, skin irritation is a known inherent risk of the device.The device manual warnings section read as follows: do not use the zio xt patch on patients with known allergic reaction to adhesives or hydrogels or with a family history of adhesive skin allergies.Patient may experience skin irritation.If skin irritation such as severe redness, itching or allergic symptoms develop, remove the zio xt patch from the patient¿s chest.This event is being reported per 21cfr803 as a serious injury.This report does not constitute an admission by rhythm that the product described in this report has any defects or has malfunctioned.These terms are included in form fda 3500a and are fixed terms for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.
 
Additional Manufacturer Narrative
This supplemental report is being submitted in response to the fda notification received on september 26, 2024, concerning missing udi numbers in mdr submissions from october 2023 through december 2024.In response to this issue, irhythm conducted an investigation and identified a system processing error, which has now been resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIO XT
Common Device Name
RECORDER, MAGNETIC TAPE, MEDICAL
Manufacturer (Section D)
IRHYTHM TECHNOLOGIES, INC
699 8th st suite 600
san francisco CA 94103
Manufacturer (Section G)
IRHYTHM TECHNOLOGIES, INC
6550 katella avenue, suite 200
cypress CA 90630
Manufacturer Contact
tara mackinnon
699 8th st suite 600
san francisco, CA 94103
6094600577
MDR Report Key20914707
Report Number3007208829-2024-00496
Device Sequence Number8466607
Product Code DSH
UDI-Device Identifier00869770000203
UDI-Public(01)00869770000203(17)250422
Combination Product (Y/N)N
PMA/510(K) Number
K121319
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date04/22/2025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2024
Supplement Date Received by Manufacturer06/02/2025
Initial Report FDA Received Date12/12/2024
Supplement Report FDA Received Date06/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
-
-