2 of 3 reports.Other mfg report numbers: 3013886523-2023-00395, 3013886523-2024-00384.A facility reported a hakim valve (id 823832), a bactiseal peritoneal catheter (id 823074) and a bactiseal ventricular catheter (id 823073) were implanted on (b)(6) 2023 and on (b)(6) 2023, the physician found out that a catheter was obstructed (customer could not confirm which catheter was obstructed); therefore the hakim valve, the peritoneal and ventricular catheters were removed on (b)(6) 2023.The patient has infectious disease - human immunodeficiency virus (hiv) with history of tubercular meningitis.The patient is stable after the revision.Infection not related to valve or catheters.
|
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The bactiseal ventricular catheter (id 823073) was not returned for evaluation as the product was discarded; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause(s) of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|