| Catalog Number |
730M0060 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that when re-introducing iliac screws intra-operatively, through force, the tips on two vital cannulated drivers became deformed.They did not fracture.There was no patient harm.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tip has twisted and fractured off.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to excessive forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference report (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that when re-introducing iliac screws intra-operatively, through force, the tips on two vital cannulated drivers became deformed.They did not fracture.There was no patient harm.This is report one of two for this event.
|
| |
|