| Catalog Number |
3301-1 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D4: the remainder of the udi number is unknown because the lot number is not available.Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore, the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2024-00390 through 3012447612-2024-00393.
|
| |
|
Event or Problem Description
|
|
It was reported that three closure tops stripped during final tightening intra-operatively.The stripped closure tops were removed and replaced.It was suspected that the torque wrench was over-torqueing and contributed to the damage.There was a 30 minute delay but no patient impact.This is report two of four for this event.
|
| |
|
Event or Problem Description
|
|
It was reported that three closure tops stripped during final tightening intra-operatively.The stripped closure tops were removed and replaced.It was suspected that the torque wrench was over-torqueing and contributed to the damage.There was a 30 minute delay but no patient impact.This is report two of four for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the device was not returned and no photos were provided, so an evaluation is unable to be performed.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to the reported torque wrench malfunction.Dhr review: the lot number was not reported and the device was not returned, therefore a dhr review is unable to be performed.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Search Alerts/Recalls
|