| Model Number |
CN LOCK SCREW, F THREAD, Ø4.0MM X 38MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/29/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On (b)(602024, it was reported by a sales representative via email that a 3024-01 11-hole left alpha plate broke postoperatively through the oblong hole in the first bend of the plate.On (b)(6) 2024, the patient was implanted with the plate, and revision surgery is still to be determined.Per the sales representative, the patient did not report any falls or injuries.No additional information has been reported.Additional information received on 11/13/2024: during the initial surgery on (b)(6) 2024, a 3024-011 11-hole left alpha plate, two 8110-026 cortical bone screw, an 8110-028 cortical bone screw, three 8110-030 cortical bone screws, an 8114-024 double lead lock cortical screw, an 8114-026 double lead lock cortical screw, an 8114-028 double lead lock cortical screw, an 8114-030 double lead lock cortical screw, an 8124-028 fully threaded cancellous locking screw, an 8124-032 cancellous locking screw, fully threaded, an 8124-038 fully threaded cancellous locking screw, an 8124-040 fully threaded cancellous locking screw, an 8124-046 fully threaded cancellous locking screw, an 8124-048 fully threaded cancellous locking screw, an 8124-050 fully threaded cancellous locking screw, an ar-8935cl-18 kreulock compression screw, and an abs-2010-10 allosync pure were implant.An abs-10062t angel cprp system with aspiration kit was also used during the procedure.On (b)(6)2024, it was discovered that the plate had broken.The patient underwent revision surgery on (b)(6)2024.All the hardware from the original surgery on (b)(6) 2024 was removed and replaced with products from another manufaturer, a 10cc of arthrex allosync pure dbm, 5cc of arthrex arthrocell plus, and bone marrow aspirate from the iliac crest.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6: one unpackaged 8124-038 cn lock screw, f thread, ø4.0mm x 38mm, serial/batch number: (b)(6), was received for evaluation.Visual inspection revealed damage to the screw threads near the head.Evaluation of the screw¿s hex head showed scratches and damage to the hex geometry.The observed damage on the screw is most likely due to it being implanted and then removed.Functional testing was conducted by screwing the returned screw into the returned 3024-011 alpha plate, left, 11-hole, serial/batch number: (b)(6).It was noted that the screw sat flush in the hole without issues.No problem was found.Arthrex concluded that there are no allegations against the 8124-038 cn lock screw, f thread, ø4.0mm x 38mm, serial/batch number: (b)(6).No related problem was found.
|
| |
|
Search Alerts/Recalls
|
|
|