Catalog Number VCP416H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a robotic inguinal hernia repair procedure in 2024 and suture was used.The suture kept coming off the suture during the case (two needles popped off during use in the case).A different new suture was used to complete the case without patient harm.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested and the following response was provided: it was mentioned that "the suture kept coming off the suture during a robotic inguinal hernia case".How many devices demonstrated the alleged deficiency? when did the needle detach or pull off from the suture (in the package, during removal from the package, during handling prior to using on the patient, or during use on the patient) ? two needles popped off during use in the case.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to fda: 1/7/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Date sent to the fda: 1/7/2025.H6 component code: g07002 no device problem found.Additional information: d9, h3, h6.H3 investigational summary: the product was returned to ethicon for evaluation.One open box with thirty-six unopened samples was received for analysis.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened.The swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.The functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 27IN 3-0 S/A SH
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key20919251
Report Number2210968-2024-13392
Device Sequence Number5108061
Product Code GAM
UDI-Device Identifier10705031053014
UDI-Public(01)10705031053014(17)290731(10)1032GL
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP416H
Device Lot Number1032GL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2024
Supplement Date Received by Manufacturer12/13/2024
Initial Report FDA Received Date12/12/2024
Supplement Report FDA Received Date01/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-