• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MOTIVA USA LLC MOTIVA IMPLANTS; ERGONOMIX ROUND SILKSURFACE WITH QID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MOTIVA USA LLC MOTIVA IMPLANTS; ERGONOMIX ROUND SILKSURFACE WITH QID Back to Search Results
Catalog Number ERSD-380Q
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/24/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As stated in the literature, ¿a number of risk factors for rupture have been identified; the most common cause is surgical instrument damage.¿ (handel, garcía and winxtrom, 2013.Breast implant rupture: causes, incidence, clinical impact, and management.Plast.Reconstr.Surg.132: 1128.) adittionally, rupture is a common and known reason for complaints, and it is clearly characterized in the product dfu included with the implant, as stated above.Additionally, no cases of rupture because of product failure have been reported to establishment labs ever.Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health.The instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: ¿breast implants can potentially remain intact for decades in the body, but all such devices will fail at some point.Breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation, but it is most likely to occur; the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress, and weakening during implantation, age and design of the implant, submuscular rather than sub-glandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture¿.Establishment labs needs to conduct the necessary tests, which include, but are not limited to, a microscopic examination and characterization of the rupture section, along with mechanical testing in compliance with approved standards.In addition, a review of the device history record was conducted, and it was concluded that there were no deviations in the manufacturing process of this device that would have caused or contibuted to this incident.When the investigation process is completed, the applicable findings will be included in a supplemental medwatch.
 
Event or Problem Description
It was reported that the right breast prosthesis demonstrated an intracapsular rupture, with no evidence of extracapsular rupture.
 
Additional Manufacturer Narrative
¿ while the product was not returned for evaluation, there was a relationship between the reported event and the device based on review of the evidence provided.A clinical evaluation of the report and evidence was executed and a rupture was confirmed.A complete review of the dhr for lot 20080358 was carried out, no deviation was found in the manufacturing process.Review of complaint history for the reported lot number and sterilization run found no other similar complaints reported in the past.¿ the instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: rupture ¿ breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation but is more likely to occur the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress and weakening during implantation, implant age and design, submuscular rather than subglandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture.Silicone gel-filled implant ruptures are most often silent; this means that most of the time, neither the doctor nor the patient can determine with the physical examination if the implant has a tear or hole in the shell.The integrity of breast implants (and detection of gel fractures and/or silent ruptures) can be evaluated through several techniques.High-resolution ultrasound (hrus) is widely accepted by healthcare providers and patients for rupture diagnosis.Additionally, the usfda recommends magnetic resonance imaging (mri) surveillance with the first mri performed three years postoperatively and subsequent mris performed every two years after that.These recommendations may vary from country to country, so please provide the patient with additional guidance based on your country's current care standards.Establishment labs does not recommend closed capsulotomy for treating capsular contracture because it can cause implant rupture.Some symptoms may appear, such as lumps surrounding the implant or in the axilla, change or loss of size or shape of the breast or implant, pain, tingling, swelling, numbness, burning, or hardening of the breast.These symptoms are not specific to rupture and may also be experienced by patients who have capsular contracture.Some cases have been reported suggesting that silicone-implant leakage should be considered in the differential diagnosis of eosinophilia.¿ while the rupture was confirmed, a definitive root cause could not be determined.Potential factors non related to the manufacture of the device that may lead to the events reported include, but are not limited to: (1) damage by surgical instruments.(2) implant stress and weakening during implantation.¿ as no manufacturing, material or design issues were identified, no further investigation or additional actions are required at this time.¿ establishment labs will continue to monitor for similar complaints/trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTIVA IMPLANTS
Common Device Name
ERGONOMIX ROUND SILKSURFACE WITH QID
Manufacturer (Section D)
MOTIVA USA LLC
125 east de la guerra street
santa barbara 93101
Manufacturer (Section G)
ESTABLISHMENTLABS S.A
building b-25, coyol free zone
alajuela,alajuela
Manufacturer Contact
susana hidalgo
125 east de la guerra street
santa barbara 93101
8003554164
MDR Report Key20919447
Report Number3012883202-2024-00216
Device Sequence Number14775830
Product Code FTR
UDI-Device Identifier07445161012203
UDI-Public(01)07445161012203(11)200821(17)250821
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/21/2025
Device Catalogue NumberERSD-380Q
Device Lot Number20080358
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/12/2024
Supplement Report FDA Received Date02/18/2025
Date Device Manufactured08/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-