• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. SMARTLOAD; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. SMARTLOAD; INTRAOCULAR LENS Back to Search Results
Model Number GCB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Endophthalmitis (1835)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section a2, a3, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section b3 - date of event: unknown/not provided.The best estimate is between implantation on (b)(6) 2024 and the awareness date 18-nov-2024.Section d6b - explant date: not applicable, as lens remains implanted.Section e1 - telephone number: (b)(6) section h3: the intraocular lens (iol) was not returned for evaluation as it remains implanted.Therefore, a failure analysis of the complaint device could not be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review and possible product return and evaluation, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that following cataract surgeries, 6 of 11 patients now receive treatment in the hospital.Bacterial infection is suspected.Patient number 6 had both eyes treated.The clinic used a preloaded monofocal intraocular lens (iol) for these patients, but the surgeon does not expect the iols to be the source of the infection.All other equipment used for the procedures were from other suppliers.Account indicated that the other 5 patients are well (patients without the infection post surgery).It was also confirmed that all surgeries took place on the same date.Overnight hospital stay and medical interventions were required.Patients have been diagnosed with endophthalmitis.No source of the infection has been found yet as per in-house investigation.There is no further patient outcome information.There is no material available for investigation.No further information provided.Note separate complaints have been created for each of the 6 patients.This report is being submitted for patient 6's left eye (os).A separate report will be submitted for this patient's right eye (od).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTLOAD
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key20920771
Report Number3012236936-2024-000321
Device Sequence Number12841953
Product Code HQL
UDI-Device Identifier05050474620643
UDI-Public(01)05050474620643(17)270522
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberGCB00
Device Catalogue NumberGCB0000220
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2024
Initial Report FDA Received Date12/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-