| Catalog Number |
MP1000 CHINA |
| Medical Device Problem Codes |
Partial Blockage (1065); Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd maxplus needleless connector was occluded the following information was received by the initial reporter with the following verbatim patient diagnosed with myelodysplastic syndromes, on (b)(6) 2024 using an indwelling needle connected to a needleless connector for infusion therapy, connecting the fluid was not smooth, multiple attempts were made to find that the needleless connector was clogged and could not be used properly, replaced with a new needleless connector, the fluid was usually, and the infusion therapy was continued without causing harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation result: a mp1000 china product was not available for investigation; however, the customer confirmed that the complaint sample was from lot 23115275.The feedback provided by the customer states that, "needle connected to a needleless connector for infusion therapy, connecting the fluid was not smooth, multiple attempts were made to find that the needleless connector was clogged and could not be used properly." no further information was available to assist the investigation in this instance.The details of this feedback were forwarded to the manufacturing site for investigation, where a review of the production records for lot 23115275 did not identify any in-process testing failures or quality deviations which may have caused or contributed to a report of this nature.The root cause of the customer¿s experience could not be determined in this instance as no sample was received for investigation.Without a sample to examine it is not possible to determine whether a manufacturing defect could have caused or contributed to a report of this nature.However, previous investigations have identified that certain designs of male luer can cause flow restriction or occlusion as they can grip onto the piston of the maxplus preventing its proper collapse.This can sometimes be resolved by disconnecting and reattaching the same luer connection which may reposition the luer against the piston and improve the flow, or alternatively by changing the connecting male luer.In this instance, without the complaint sample to examine it has not been possible to conclusively link this feedback to a specific failure mode; however, a review of the customer feedback database indicates that this is a rare occurrence with a small number of similar reports against the product mp1000 china in the past 12 months.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|