| Catalog Number |
8065752181 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that aspiration was weak during cataract (with intraocular lens implant) surgery.The surgery was completed after replacing the pak with another one.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in d.9., h.3., h.6.And h.11.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.Current tracking indicates no adverse trend for this lot for this event.The returned sample was visually inspected and functionally tested.Both irrigation and aspiration luers were intact.Flare shape was observed in the blue socket fitting at the aspiration tubing insertion end.The sample was tested using a calibrated centurion console with the current software revision.The sample could be primed and tuned, and the service data could be retrieved from the console.The aspiration and irrigation free flows met specification.The root cause of the customer's complaint could not be established since the returned sample met specifications.No contributing factors could be identified that could cause the reported complaint.After the investigation of this complaint, it has been determined that this sample met specifications.Therefore, no action will be taken at this time.Based on our current tracking, there are no adverse trends for this reported complaint.Quality assurance will continue to monitor customer complaints via the complaint review meetings and will take action for any future occurrences as is deemed necessary.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|