| Model Number |
N/A |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during inspection of stock, the sterile pouch was found damaged.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign: country: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Updated: d9, g3, g6, h2, h3, h6, h10.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (pouch).Sterility has been compromised.Additional evaluation found damage to the cartons; however, as this damage was not initially reported, no further investigation was performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.The reported event has been confirmed by evaluation of the returned product and provided photos.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|
Search Alerts/Recalls
|