Catalog Number 180562
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial shoulder arthroplasty with non-zimmer biomet products on and unknown date.The patient then underwent a revision surgery approximately one (1) year ago due to an unknown reason.Subsequently, several months after the revision surgery, there appeared to be loosening of the glenoid implants.The patient will undergo a revision surgery on an unknown date.
 
Additional Manufacturer Narrative
(b)(4).D10: medical products: item#: 010000589, comp rvrs 25mm bsplt ha+adptr; lot#: 66053299; item#: 115397, comp rvs cntrl 6.5x35mm st/rst; lot#: 66082758; item#: 180559, comp nlk scr 3.5hex 4.75x25 st; lot#: 66127676; item#: 180562, comp nlk scr 3.5hex 4.75x40 st; lot#: 859790; item#: 180561, comp nlk scr 3.5hex 4.75x35 st; lot#: 66138820; item#: 180559, comp nlk scr 3.5hex 4.75x25 st; lot#: 66127676; item#: 115316, comp rvrs shldr glnsp +6 36mm; lot#: 135990; item#: 00-4349-010-17, tm reverse stem 10mm x 170mm; lot#: 65550644; item#: 00-4349-036-00, 36mm poly liner plus 0mm ofs; lot#: 65871436.G2: foreign: france.H10: 0001825034-2024-02361-1, 0001825034-2024-02362-1, 0001825034-2024-02363-1, 0001825034-2024-02364-1, 0001825034-2024-02365-1, 0001825034-2024-02366-1.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: reverse total shoulder arthroplasty with fracture glenoid screws but no dislocation.Sizing is appropriate.There are multiple fractured glenoid screws from a reverse total shoulder arthroplasty.It was noted that 5 peripheral screws were invoiced per usage and sticker sheets.The baseplate only has openings for 4.However, it is unknown which item was not implanted and why it was not used.As such, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP NLK SCR 3.5HEX 4.75X40 ST
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20924832
Report Number0001825034-2024-02941
Device Sequence Number20100795
Product Code PHX
UDI-Device Identifier00880304677296
UDI-Public(01)00880304677296(17)320808(10)919680
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K130390
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number180562
Device Lot Number919680
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2024
Initial Report FDA Received Date12/13/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-