Corrected information: h3.Yes h6.Investigation findings: 3252 investigation conclusions: 61 device evaluation: visual inspection confirms that the hexalobe driver tip has broken off.The tip likely sheared off from torsional overload.A review of device history records showed no manufacturing or processing related irregularities.Labeling review: warnings/cautions/precautions: risks identified with the use of these devices, which may require additional surgery, include device component failure, loss of fixation/stabilization, non-union, vertebral fracture, neurological injury, vascular or visceral injury.Possible adverse effects: initial or delayed loosening, disassembly, bending, dislocation, and/or breakage of device components.
|