Corrected information: h3.Yes.H6.Investigation findings: 3252.Investigation conclusions: 61.Device evaluation: visual inspection confirms that the majority of the hexalobe driver tip has broken off.This is likely a result of improper loading before torque delivery through the driver tip.A review of device history records showed no manufacturing or processing related irregularities.Labeling review: warnings/cautions/precautions: risks identified with the use of these devices, which may require additional surgery, include device component failure, loss of fixation/stabilization, non-union, vertebral fracture, neurological injury, vascular or visceral injury.Possible adverse effects: initial or delayed loosening, disassembly, bending, dislocation, and/or breakage of device components.
|