EXACTECH, INC. ADVANCED PATELLA 32MM 3 PEG IMPLANT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
200-07-32 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Scar Tissue (2060); Synovitis (2094); Joint Laxity (4526); Swelling/ Edema (4577)
|
| Date of Event |
03/31/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(6) -2023-02960 d10: concomitants: (b)(6) 02-022-45-3030 - truliant tib fit tray cem sz 3f / 3t.(b)(6) 02-020-11-0230 - truliant ps cem fem ps cem left sz 3.(b)(6) 02-012-60-1425 - tru stem ext 14mm x 25mm.(b)(6) 204-70-00 - tibial stem ext.Screw.(b)(6) 200-07-32 - advanced patella 32mm 3 peg implant.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-02960.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, a failure of the cement used to secure the femoral component to the femoral bone, which led to loosening at the cement-implant interface, and/or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 54 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, loosening, instability, synovitis, joint laxity, scar tissue, failure of implant, swelling/edema, and pain.Initial surgery and revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.
|
| |
|
Search Alerts/Recalls
|
|
|