|
H8 - usage of device: correction.Manufacturer's investigation conclusion: the reported event of yellow wrench and red battery indicator issue could not be confirmed with the returned 12v sealed lead acid battery (serial # (b)(6)).The 12v battery was installed into a test power module and an expected yellow battery charging indicator was active.The battery completed the charge process when the battery charging indicator turned green.The system was able to operate on the battery until a red battery alarm became active approximately 67 minutes later.The battery pass specification, which is greater than 60 minutes.The battery functioned as intended and the reported yellow wrench and red battery indicator issue could not be reproduced.A root cause of the reported yellow wrench and red battery indicator issue could not be determined through this evaluation.The complaint does not meet the requirement of the investigation procedure for a review of the device history records.Heartmate 3 instructions for use has details on the proper use of the power module.If the power module does not function properly, contact the company's technical service department for assistance.Under section 3, ¿powering the system¿, describes all audible and visual alarms.Also, under this section, indicator symbols on the front panel of the power module illuminate to indicate the charge status of the power module backup battery.Table 3.1 describes the indicator symbols.To set up the power module, perform these tasks: install the power module backup battery.Connect the power cord.Connect the power module patient cable.In section f, safety checklists, replace any equipment or system component that appears damaged or worn.¿yearly safety checklist¿, schedule a power module inspection and cleaning with thoratec-trained personnel.The inspection and cleaning includes (but is not limited to) functional testing, cleaning, inspection, and replacement of the power module backup battery.No further information was provided.The manufacturer is closing the file on this event.
|