| Model Number |
C4130 |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
Physical Entrapment (2327)
|
| Date of Event |
07/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event unit returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of adhesion at the front tip as multiple frays were observed on the latis pads.Based on the condition of the returned unit and the description of the event, it is likely that the reported event was caused by frays identified on the latis pad.Applied medical has performed a historical trend analysis and review of production records and no trends were identified.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.Based upon the initial description of the event, the event was not reportable as it is unlikely to cause or contribute to death or serious injury.However, based on the evaluation of the returned unit and further examination of the event description, applied medical determined that this event is reportable.This report represents the initial and follow-up report.
|
| |
|
Event or Problem Description
|
|
Procedure performed: robotic lar.Event description: complaint 1 of 2: (b)(4), complaint 2 of 2: (b)(4).[hospital] after using this instrument for a period of time, the front tip will become adhesive to tissues.Changing to another instrument for use will still result in adhesion at the front tip, until the third grasper is used to complete the surgery.There is no impact to patient.User replace with new grasper to complete this surgery.There is no other instruments to affect this event.Product is available for return.Additional information received on 15aug2023 via email from [distributor]: intestine tissue was being grasped at the time of the incident.Intervention: user replace with new grasper to complete this surgery.Patient status: fine.
|
| |
|
Search Alerts/Recalls
|