EXACTECH, INC. 145-DEG PE 42MM HUM LINER +0; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-42-00
Medical Device Problem Codes Detachment of Device or Device Component (2907); Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/25/2024
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 5 years post initial left reverse total shoulder arthroplasty (tsa), a revision was performed due to polyethylene liner dissociation.The surgeon removed a 42 glenosphere, humeral tray, liner and screws and replaced with a 42+4 glenosphere, +0 humeral tray, glenosphere locking screw, reverse shoulder torque screw kit and new 42+0 liner.The event was unrelated to the breakage of a device and did not lead to a surgical delay/prolongation.The patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
Customer has indicated that the product is in process of being returned for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: d1, d4, g4, h6.Mdr section codes updated/corrected: b, c, d.The revision reported was likely the result of disassembly of the humeral liner from the humeral tray and unintended articulation of metal components likely resulted in metal-on-metal wear and subsequent metallosis.However, this cannot be confirmed because the components were not returned for evaluation.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
145-DEG PE 42MM HUM LINER +0
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct.
gainesville, FL 32653
3523771140
MDR Report Key20940774
Report Number1038671-2024-04847
Device Sequence Number18193286
Product Code KWS
UDI-Device Identifier10885862086693
UDI-Public10885862086693
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number320-42-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2024
Supplement Date Received by Manufacturer10/31/2025
Initial Report FDA Received Date12/16/2024
Supplement Report FDA Received Date11/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EQUINOXE HUMERAL TRAY
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexMale
-
-