| Catalog Number |
MQ1816 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records was performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to an ultrasound guided breast biopsy procedure through normal tissue density, the device label was allegedly mislabeled.No coaxial was used and the procedure was completed using another device.There was no patient contact.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to an ultrasound guided breast biopsy through normal density tissue, the device label was allegedly mislabeled.No coaxial was used and the procedure was completed using another device.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported labelling issue could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported labelling issue could not be determined based upon the provided information.Labeling review: labeling review is required for this record as the reported event is a labeling issue.By comparing the received carton label sticker from the manufacturer site against the carton label document, it was noted that everything mentioned in the product label sticker such as pk number, country origin, gauge size & needle length, lot number characters, use by date format, gtin number, sequential number format was verified and it matched with the document.By comparing the received tylek lid label sticker from the manufacturer site against tylek lid document, it was noted that the catalogue number, pk number, contents, gauge size & needle length, lot number characters, use by date format, gtin number which was mentioned in the tylek lid was verified and it matched with the document.By comparing the first and last carton and tylek lid labels from the manufacturer¿s dhr with specific documents, it was confirmed that all fields were consistent with the specification requirements.B5, g1, g2, g3.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, a patient underwent an ultrasound guided breast biopsy procedure through normal tissue density.It was reported that prior to the procedure the device label was allegedly mislabeled.No coaxial was used and the procedure was completed using another device.There was no patient contact.
|
| |
|