| Model Number |
36010 |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: initial reporter phone no (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a spectrum v9 wireless battery module over heated.This occurred during testing in the biomedical service department.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, h3, h6, h11.H11: the device was received for evaluation.Functional testing was performed and did not identify any issues related to the customer reported condition.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|