| Catalog Number |
1788010000 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: loss of image.Confirmed failure: dx codes verbiage: software reprogram software upgrade.Probable root cause: faulty solder joints in video path.Faulty prism board or connectors.Faulty xboard or ch cable connectors.Break in ch cable core.Faulty hdmi cable.Faulty ccu power supply.Internal ccu cable failure - power and/or communication.Improper/no fuse installed.Ac inlet board.Main board - video processor.Digital board - fpga.Transition board - coax connector.Electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge.Poor system grounding.Improper ccu timing.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|