| Catalog Number |
1788010000 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: loss of image confirmed failure: dx codes verbiage: software reprogram softwar upgrade probable root cause: - faulty solder joints in video path - faulty prism board or connectors - faulty xboard or ch cable connectors - break in ch cable core - faulty hdmi cable - faulty ccu power supply - internal ccu cable failure - power and/or communication - improper/no fuse installed - ac inlet board - main board - video processor - digital board - fpga - transition board - coax connector - electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge - poor system grounding - improper ccu timing.
|
| |
|