| Catalog Number |
71674084 |
| Medical Device Problem Codes |
Fracture (1260); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during an internal fixation surgery for a hip fracture, the intertan nail cap 0.0mm was stripped (case (b)(4)), it got stuck to a (1) medium hexdriver (case (b)(4)).Additionally, the end of a (1) screwdriver release handle was bent and broke off while trying to get the capture off the medium hex driver (case (b)(4)).After removing the nail cap, it broke off on the back table.The procedure was resumed, without any delay, using a s+n back-up device.No injury was reported as a consequence of this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an internal fixation surgery for a hip fracture, the end of a (1) screwdriver release handle was bent and broke off while trying to get the capture off the medium hex driver.The procedure was resumed, without any delay, using a s+n back-up device.No injury was reported as a consequence of this issue.Patient's status is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device was returned and evaluated.The visual inspection revealed that the tip of the device is fractured off.The remaining parts of the end of the device is bent and twisted.The device shows signs of wear from use.This inspection was conducted by the naked eye.A review of the production orders did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed part number over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|