SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HEAD 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
|
Back to Search Results |
|
| Catalog Number |
74121150 |
| Medical Device Problem Code |
Biocompatibility (2886)
|
| Health Effect - Clinical Codes |
Scar Tissue (2060); High metal ion levels (4847)
|
| Date of Event |
08/15/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, a plaintiff received a right bhr system on (b)(6) 2020.The bhr failed and patient experienced pain and tested for elevated cobalt chromium ions, since (b)(6) 2023, therefore a revision surgery was performed on (b)(6) 2024.During surgery the acetabular cup was found to be well seated and was kept in place.The hip was converted into a tha using s+n components.Patient was transferred to the recovery room in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
H10.Additional information in d10.H11.Corrected information in h6 (health effect - clinical code).
|
| |
|
Additional Manufacturer Narrative
|
|
H3,h6: as of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the alleged devices was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for the head and cup.This failure will continue to be monitored via routine trending.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.The review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, no prior applicable escalation actions were identified.No further escalation actions are required.The available medical documents were reviewed.It should be noted, the bhr surgical technique indicates, the acetabular component is to be impacted with 15-20° of anteversion and 40-45° inclination angle (bhr ifu 04727 v2 45670103 revc 01/18).A clinical root cause of the elevated metal ions cannot be confirmed.The scar tissue cannot be ruled out as a possible contributing factor to the reported pain.Scar tissue is a known complication of joint surgery, and it cannot be concluded to be related to the implant.The patient impact beyond the reported revision could not be determined.Based on the information provided, further investigation of the reported complaint cannot be carried out and remains inconclusive.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
|
| |
|