| Catalog Number |
0279351400 |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device had insufficient / no energy delivered.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: serfas not working probable root cause: probe short (possibly due to fluid ingress), open circuit (possibly due to loose elecrical connection), power output variation, wrong default setting, nonconforming hand piece cable console error crossflow temperature mitigation actions use error the reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the device had insufficient / no energy delivered.
|
| |
|
Search Alerts/Recalls
|