|
|
| Catalog Number |
72201781 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an acl reconstruction surgery, the biosure ha screw was fractured while it was twist in.The procedure was completed using a back-up device.There was a delay greater than 30 minutes and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the reported device was received for evaluation.A visual inspection revealed it is not in its original packaging.The device was returned with the distal end fractured away, there is biological debris on the returned fragment.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use).Based on this investigation, there is no evidence to suggest a design, material or manufacturing issue.Therefore, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|
|
|