MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507550BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Deformity/ Disfigurement (2360); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no manufacturing record evaluation could be performed.Reason for device explant and/or reoperation: silent rupture, chest injury.H6 health effect - clinical code e2402: chest injury.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female who underwent a breast augmentation revision with an unknown silicone prosthesis experienced bilateral chest injury, and bilateral silent ruptures postoperative.The patient had a scan a few months back that showed ruptured implants.Possible trauma due by mammogram examination.As a result, patient underwent bilateral removal on (b)(6) 2024.This report relates to the left side.
|
| |
|
Additional Manufacturer Narrative
|
|
Suspect device received on january 11, 2025.Device evaluation completed on january 16, 2025: the product was returned to mentor for evaluation.Mentor conducted the visual inspection.Visual analysis of the returned sample revealed that the gel mod-rnd 550cc breast implant was ruptured.In addition, shell abrasion was observed at the edges of the rupture.The evaluation determined that the possible cause of the rupture is the trauma experienced.Rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Trauma is a known inherent risk of gel-filled mammary prosthesis.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.The manufacturing record evaluation (mre) was reviewed, and no anomalies were found related to this complaint.In addition, the mre review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|