• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE Back to Search Results
Catalog Number 53616G
Medical Device Problem Code Incomplete or Missing Packaging (2312)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that intermittent catheters were leaking.Per customer follow up received on 03dec2024, it was reported that they thought cardinal misreported this incident - the issue was that the adhesive tab on the packaging was missing, and the patient relies on that packaging for use.After discussing the issue with their representative, they found that there was a brief period of time where the adhesive tab was removed.
 
Event or Problem Description
It was reported that intermittent catheters were leaking.Per customer follow up received on 03dec2024, it was reported that they thought cardinal misreported this incident - the issue was that the adhesive tab on the packaging was missing, and the patient relies on that packaging for use.After discussing the issue with their representative, they found that there was a brief period of time where the adhesive tab was removed.
 
Additional Manufacturer Narrative
Upon further review, bd has determined that this event is not reportable.We are providing this record as additional information.The reported event was inconclusive because no sample was available for evaluation and further investigation did not result in any additional findings.No actions can be taken at this time since a root cause was not identified.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.A labeling review was not performed because labeling could not have prevented the reported failure.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE
Common Device Name
INTERMITTENT CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key20986400
Report Number1018233-2024-08141
Device Sequence Number8580847
Product Code KOD
UDI-Device Identifier00801741069789
UDI-Public(01)00801741069789(11)240717(17)290528(10)JUJT1029
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K000723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number53616G
Device Lot NumberJUJT1029
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2024
Supplement Date Received by Manufacturer01/13/2025
Initial Report FDA Received Date12/20/2024
Supplement Report FDA Received Date01/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-