• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH SIMPLEX HV US 1 PACK; BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH SIMPLEX HV US 1 PACK; BONE CEMENT Back to Search Results
Catalog Number 6194-1-001
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 11/19/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Pt.Called to report that they had a procedure on (b)(6) 2023 with stryker implants on their left knee.They stated that they experienced loosening and some of the implant failed.They would like to file a claim.Additional info.27-nov-2024: revision on (b)(6) 2023.Dec 20.2024- as per medical review, the patient was revised due to tibial loosening and rom.An additional record was created fort he mua after revision.
 
Additional Manufacturer Narrative
Reported event: an event regarding loosening involving a simplex cement mix was reported.The event was confirmed via medical review.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of medical records by clinician indicated "patient underwent a tka with the robot.The procedure appeared routine.The decision for uni vs total was made in the or and a decision to move to total based on the trochlear cartilage.No preoperative records or radiographs were provided for review.The postoperative course and complete x-rays were not provided for review.The patient presented for a second opinion, the reasons appeared to be dissatisfaction with the knee¿s rom and some pain.In the second opinion it was stated that there was a proposal for a debridement and poly exchange as a treatment course.The second opinion did order hip x-rays and an mri to evaluate radicular like pan in the left lower extremity.The knee evaluation was not particularly specific.After checking inflammatory markers, it was felt that revision was in order for a loose tibia and stiffness, the description of x-rays was simply loosening along the tibia.An evaluation of the loosening was not given.The revision operative note was not provided.Postoperatively from the revision the patient underwent mua for stiffness.The course thereafter was not provided for review.Event confirmation: a revision surgery of a mako tka was performed.An mua was also performed on the revision.Two reasons for the revision were given the first, stiffness, can be confirmed from the record.The second, tibial loosening, cannot be confirmed.Root cause: the root cause of the revision was stiffness and presumed tibial loosening.The tibial loosening could not be confirmed and thus no root cause can be attributed.The root cause of the stiffness cannot be ascertained." -product history review: a review of the device history records was not performed as the device is oem.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to loosening."patient underwent a tka with the robot.The procedure appeared routine.The decision for uni vs total was made in the or and a decision to move to total based on the trochlear cartilage.No preoperative records or radiographs were provided for review.The postoperative course and complete x-rays were not provided for review.The patient presented for a second opinion, the reasons appeared to be dissatisfaction with the knee¿s rom and some pain.In the second opinion it was stated that there was a proposal for a debridement and poly exchange as a treatment course.The second opinion did order hip x-rays and an mri to evaluate radicular like pan in the left lower extremity.The knee evaluation was not particularly specific.After checking inflammatory markers, it was felt that revision was in order for a loose tibia and stiffness, thee description of x-rays was simply loosening along the tibia.An evaluation of the loosening was not given.The revision operative note was not provided.Postoperatively from the revision the patient underwent mua for stiffness.The course thereafter was not provided for review.Event confirmation: a revision surgery of a mako tka was performed.An mua was also performed on the revision.Two reasons for the revision were given the first, stiffness, can be confirmed from the record.The second, tibial loosening, cannot be confirmed.Root cause: the root cause of the revision was stiffness and presumed tibial loosening.The tibial loosening could not be confirmed and thus no root cause can be attributed.The root cause of the stiffness cannot be ascertained." additionally, the reported device is an oem product.Written acknowledgement from the oem supplier confirmed that they have done retain tests and found no deviations.See supplier issue ref #rm2024/146.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Pt.Called to report that they had a procedure on (b)(6) 2023 with stryker implants on their left knee.They stated that they experienced loosening and some of the implant failed.They would like to file a claim.Additional info.27-nov-2024: revision on (b)(6) 2023.Dec 20.2024- as per medical review, the patient was revised due to tibial loosening and rom.An additional record was created fort he mua after revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLEX HV US 1 PACK
Common Device Name
BONE CEMENT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key20999381
Report Number0002249697-2024-01782
Device Sequence Number21373038
Product Code LOD
UDI-Device Identifier04260056885747
UDI-Public04260056885747
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number6194-1-001
Device Lot Number211AC9621C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/02/2024
Supplement Date Received by Manufacturer03/31/2025
Initial Report FDA Received Date12/23/2024
Supplement Report FDA Received Date04/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexMale
Patient Weight190 KG
Patient EthnicityHispanic
-
-