BIOMET 3I OSSEOTITE 2 IMPLANT 5 X 8.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number XFOS585
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Nerve Damage (1979)
Date of Event 09/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor reports that the implant had to be removed due to nerve damage.Procedure not completed.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).Zimvie received one (1) xfos585, (sseotite 2 implant 5 x 8.5mm) for evaluation.Visual evaluation was performed, the returned implant has signs of use, apparent bone / tissue attached to external threads.The implant was identified as reported however, no apparent malfunction was identified with the implant that would contribute to the reported event.Implant matched prints where measured.Device history record (dhr) review was completed for the subject lot number 2018080690.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2018080690 for similar events and no other complaint was identified.Review completed utilizing keywords: dental: medical: nerve damage.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was customer error.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE 2 IMPLANT 5 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key20999702
Report Number0001038806-2024-02988
Device Sequence Number9961278
Product Code DZE
UDI-Device Identifier00844868023944
UDI-Public(01)00844868023944(17)230913(10)2018080690
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K111216
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 03/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date09/12/2023
Device Catalogue NumberXFOS585
Device Lot Number2018080690
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/23/2024
Supplement Report FDA Received Date03/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/12/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient EthnicityNon Hispanic
-
-