| Catalog Number |
CDS0701-XTW |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.An xtw clip was prepared per the instructions for use (ifu) and inserted without issues.However, while in the valve, it was observed that one gripper was not functioning as intended.Troubleshooting was performed, but the issue was unable to be resolved.Therefore, the clip was removed and replaced.Two clips were then deployed, reducing mr to a grade of 1.There were no adverse patient effects and no clinically significant delay in the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned device analysis did not confirm the reported single gripper actuation issue (difficult to open or close).A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.A review of the complaint history identified no other complaints from the lot.Based on information provided and the results of the returned device analysis (unable to confirm the reported issue), a cause for the reported difficult to open or close associated with a single gripper actuation issue could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.An xtw clip was prepared per the instructions for use (ifu) and inserted without issues.However, while in the valve, it was observed that one gripper was not functioning as intended.Troubleshooting was performed, but the issue was unable to be resolved.Therefore, the clip was removed and replaced.Two clips were then deployed, reducing mr to a grade of 1.There were no adverse patient effects and no clinically significant delay in the procedure.
|
| |
|