| Catalog Number |
A800361 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a sterile water product about to be used for a foley catheter and it was found to be contaminated.Per additional information via email on 14jan2025, it was reported that the category of contamination was unknown.A staff member noticed a foreign substance floating in the sterile water fluid.This occurred on 22sep2024.They unfortunately did not have the faulty product, therefore, no physical sample is available.There was no patient involvement, this was fortunately caught before administration to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed cause unknown.No actions can be taken at this time since a root cause was not identified.Visual evaluation of the returned photo sample noted one opened (with original packaging), foley tray component sterile water (syringe).Visual inspection of the photo sample noted foreign material inside the sterile water syringe.This does not meet specification which states loose foreign matter.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: patient has acute urinary retention or bladder outlet obstruction need for accurate urine output measurements in critically ill patients use for selected surgical procedures to assist in healing of open sacral or perineal wounds patient requires prolonged immobilization to improve comfort for end of life care intended for use in the drainage and/or collection and/or measurement of urine.Generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder.However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as a nephrostomy tract.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a sterile water product about to be used for a foley catheter and it was found to be contaminated.
|
| |
|
Search Alerts/Recalls
|