AVANOS MEDICAL INC. BALLARD TURBO-CLEANING CLOSED SUCTION SYSTEM; TURBO-CLEANING CLOSED SUCTION CATHETER, T-PIECE Back to Search Results
Model Number 99734
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is reported to be available but has not yet been received by the manufacturer.A review of the device history record is in-progress all information reasonably known as of 24 dec 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
 
Event or Problem Description
Sunmed holdings llc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the first of two reports.Refer to 8030647-2024-00097 for the second report.It was reported, the tube broke/ snapped at the on/off button end; there was no reported injury.
 
Additional Manufacturer Narrative
Additional information: d4.Correction: g1.The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 11 apr 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
 
Event or Problem Description
Sunmed holdings llc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the first of two reports.Refer to 8030647-2024-00097 for the second report.It was reported, the tube broke/ snapped at the on/off button end; there was no reported injury.
 
Additional Manufacturer Narrative
Additional information: d4; h6.The actual sample, from the reported event was returned and evaluated; the reported issue was not duplicated during evaluation.The reporter provided photographic/video evidence which depicted the reported issue and the complaint was confirmed.However; as the reported failure was not reproduced in the lab during sample evaluation the root cause could not be determined.The device history record for lot 30326532 was reviewed and the product was produced according to product specifications.All information reasonably known as of 09 may 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
 
Event or Problem Description
Sunmed holdings llc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the first of two reports.Refer to 8030647-2024-00097 for the second report.It was reported, the tube broke/ snapped at the on/off button end; there was no reported injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLARD TURBO-CLEANING CLOSED SUCTION SYSTEM
Common Device Name
TURBO-CLEANING CLOSED SUCTION CATHETER, T-PIECE
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, mx 84160
MX   84160
Manufacturer Contact
racheal cooper
2710 north ridge dr nw
ste a
grand rapids, MI 49544
2054826767
MDR Report Key21014411
Report Number8030647-2024-00096
Device Sequence Number9052202
Product Code BSY
UDI-Device Identifier00609038997346
UDI-Public(01)00609038997346(10)30326532(17)290514(240)120002563
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number99734
Device Catalogue Number22716-5
Device Lot Number30326532
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2024
Supplement Date Received by Manufacturer03/17/2025
04/17/2025
Initial Report FDA Received Date12/24/2024
Supplement Report FDA Received Date04/11/2025
05/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-