MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC EM ENT STRAIGHT SUCTION; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC EM ENT STRAIGHT SUCTION; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9733449
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2023
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding an instrument during a functional endoscopic sinus surgery (fess) procedure.It was reported that the straight suction would not verify for a case.The site ended up not using the instrument in a nav capacity for the case.There was less than an hour delay in surgery.There was no impact on patient outcome.
 
Additional Manufacturer Narrative
H3, h6) the returned suction had difficulty verifying when attached to a known good tracker.This regulatory report is being submitted due to retrospective review through capa 626390.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM ENT STRAIGHT SUCTION
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key21019765
Report Number1723170-2024-04037
Device Sequence Number14483319
Product Code HAW
UDI-Device Identifier00763000053369
UDI-Public00763000053369
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9733449
Device Catalogue Number9733449
Device Lot Number200526
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2023
Initial Report FDA Received Date12/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age34 YR
Patient SexMale
-
-