• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Catalog Number 3502175
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 12/09/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient who underwent a primary breast reconstruction procedure with mentor smooth round moderate plus profile 175cc saline breast prostheses developed baker grade iii-iv capsular contracture in both breasts post implantation.As a result, the patient underwent bilateral explantation and replacement with unspecified mentor gel breast prostheses on (b)(6) 2024.This medwatch report is for the patient¿s right breast prosthesis.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 14-jan-2025, the mentor failure analysis lab received the device for evaluation.On 20-jan-2025, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information reported, the patient developed capsular contracture.The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet the american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsules, and effusion to pathology for examination.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND MODERATE PLUS PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key21020681
Report Number1645337-2024-15011
Device Sequence Number3969251
Product Code FWM
UDI-Device Identifier00081317001348
UDI-Public(01)00081317001348(17)181024(10)6876793
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/24/2018
Device Catalogue Number3502175
Device Lot Number6876793
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/09/2024
Supplement Date Received by Manufacturer01/14/2025
Initial Report FDA Received Date12/26/2024
Supplement Report FDA Received Date01/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-