| Model Number |
10.4.1 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Improper or Incorrect Procedure or Method (2017); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Merge healthcare has performed a root cause investigation of the heart rate issue reported by the user facility.The invesitgation determined that the merge hemo system is working in accordance with its design and is meeting specified performance parameters.Merge healthcare is working with the customer to resolve external factors in the use environment that are affecting performance of the system.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, merge healthcare received a copy of a medsun report.A merge hemo customer reported: heart rate reporting as 4 beat per minute (bpm), actual heart rate 70 bpm.Merge healthcare has requested additional information from the customer for this event.There have been no reports of patient injury or harm because of this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Merge healthcare has performed a comprehensive investigation and risk to health analysis of the reported issue.The investigation determined that no malfunction of the device occurred, and the health hazard evaluation concluded that the event is unlikely to result in any adverse health consequences.No patient harm has occurred as a result of the events reported by the customer facility.Merge healthcare will continue to follow up with facility staff to prevent future occurrences of the issue.Revised information contained in this supplemental report include the following: g6 - indication that this is follow-up report 001.H2- indication of additional information.H6 - investigation conclusion code: 4323.This is the final report for 2183926-2024-00027.
|
| |
|