MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED2-450-20
Medical Device Problem Codes Retraction Problem (1536); Material Deformation (2976); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information that a ped2 pipeline stent had resistance in the catheter.The doctor found it extremely difficult to deliver the stent into the plastic microcatheter, and the plastic microcatheter was unable to retract the stent, resulting in both the stent and the plastic microcatheter being unusable.The devices were prepared as indicated in the instructions for use (ifu).The patient was being treated for an unruptured, saccular aneurysm in the ophthalmic artery.The max diameter was 4mm and the neck diameter was 5mm.The distal landing zone was 8mm and the proximal landing zone was 10mm.Vessel tortuosity was normal.Dual antiplatelet treatment was administered.Angiographic result post procedure was good.No patient symptoms or complications were reported.Ancillary devices: medtronic guidewire the resistance was in the distal end of the catheter.It was observed that the device looked "rough".
 
Event or Problem Description
Additional information was received that the pushwire looked "rough".
 
Additional Manufacturer Narrative
H3: product analysis ¿as found condition: the pipeline flex shield embolization device was returned for analysis within shipping box; and within an opened pipeline flex shield inner pouch.¿damage location details: no bend was observed on the pushwire.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.No defects were found with the tip coil, distal marker, re-sheathing marker, pads or with the proximal bumper.The braid was returned already detached from the pusher; therefore, the proximal and distal ends could not be identified.The pipeline flex shield braid ends appeared to be fully opened and damaged.No other anomalies were observed.¿ conclusion: based on the analysis findings, the pipeline flex shield could not be confirmed to have resistance as no defect was found with pushwire.In addition, the micro catheter used in this event was not returned.Therefore, any contributing factors could not be assessed.However, the root cause could not be determined.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX W/SHIELD TECHNOLOGY
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key21021740
Report Number2029214-2024-02435
Device Sequence Number19825838
Product Code OUT
UDI-Device Identifier00763000284541
UDI-Public00763000284541
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P100018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPED2-450-20
Device Catalogue NumberPED2-450-20
Device Lot NumberB678514
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2024
Supplement Date Received by Manufacturer01/15/2025
Initial Report FDA Received Date12/26/2024
Supplement Report FDA Received Date01/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-