| Model Number |
PED2-450-20 |
| Medical Device Problem Codes |
Retraction Problem (1536); Material Deformation (2976); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that a ped2 pipeline stent had resistance in the catheter.The doctor found it extremely difficult to deliver the stent into the plastic microcatheter, and the plastic microcatheter was unable to retract the stent, resulting in both the stent and the plastic microcatheter being unusable.The devices were prepared as indicated in the instructions for use (ifu).The patient was being treated for an unruptured, saccular aneurysm in the ophthalmic artery.The max diameter was 4mm and the neck diameter was 5mm.The distal landing zone was 8mm and the proximal landing zone was 10mm.Vessel tortuosity was normal.Dual antiplatelet treatment was administered.Angiographic result post procedure was good.No patient symptoms or complications were reported.Ancillary devices: medtronic guidewire the resistance was in the distal end of the catheter.It was observed that the device looked "rough".
|
| |
|
Event or Problem Description
|
|
Additional information was received that the pushwire looked "rough".
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis ¿as found condition: the pipeline flex shield embolization device was returned for analysis within shipping box; and within an opened pipeline flex shield inner pouch.¿damage location details: no bend was observed on the pushwire.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.No defects were found with the tip coil, distal marker, re-sheathing marker, pads or with the proximal bumper.The braid was returned already detached from the pusher; therefore, the proximal and distal ends could not be identified.The pipeline flex shield braid ends appeared to be fully opened and damaged.No other anomalies were observed.¿ conclusion: based on the analysis findings, the pipeline flex shield could not be confirmed to have resistance as no defect was found with pushwire.In addition, the micro catheter used in this event was not returned.Therefore, any contributing factors could not be assessed.However, the root cause could not be determined.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|