AOMORI OLYMPUS CO., LTD. BML HANDLE V Back to Search Results
Model Number MAJ-441
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a therapeutic ercp (endoscopic retrograde cholangiopancreatography) under sedation, an attempt was made to release impaction using the handle, but it could not be released.The guidewire of the single use mechanical lithotripter was tangled in the basked.After the basket was removed, the enbd was placed in the bile duct and the procedure was completed.Due to the device malfunctions, the procedure was prolonged by 45 minutes.There were no reports of patient harm.
 
Additional Manufacturer Narrative
Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.Olympus will continue to monitor the field performance of this device.This report is related to (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BML HANDLE V
Common Device Name
HANDLE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key21022209
Report Number9614641-2024-20135
Device Sequence Number9207487
Product Code LQC
UDI-Device Identifier04953170218446
UDI-Public04953170218446
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 12/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAJ-441
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/27/2024
Initial Report FDA Received Date12/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BML-V437QR-30
Patient SexUnknown
-
-